BRT AA1800 dedicated flame AAS with double-beam optics, 185–rango de 900 nm, eight-lamp turret and titanium burner head. Full datasheet specifications.
Product overview
El BRT AA1800 is a dedicated flame atomic absorption spectrophotometer. It is built for trace and major element determination in environmental monitoring, food safety testing and metallurgical analysis, using a double-beam optical path, an eight-lamp turret and a titanium burner head.

Key features across the series
- Total-reflection optical system designed for high signal-to-noise performance.
- Eight-lamp turret with 360° rotation and simultaneous preheating of 1–4 lamps.
- Windows workstation software with automatic control, QA/QC functions, SD/RSD calculation and calibration-curve correlation checking.
- Multiple data-processing modes including average, peak height, peak area and statistical analysis.
- One-click database access for wavelength, ancho de banda espectral, lamp current, negative high voltage and energy-balance settings.
Optical system and photometric performance
| Parámetro | AA1800 | AA1900 | AA2000 |
|---|---|---|---|
| Instrument type | Flame system | Flame & GFAA integrated | Flame & GFAA integrated |
| Beam type | Double-beam | Double-beam | Double-beam |
| Diffraction grating | 1800 lines/mm; blaze wavelength 250 Nuevo Méjico | 1800 lines/mm; blaze wavelength 250 Nuevo Méjico | 1800 lines/mm; blaze wavelength 250 Nuevo Méjico |
| Rango de longitud de onda | 185–900 nm | 185–900 nm | 185–900 nm |
| Ancho de banda espectral | 0.1, 0.2, 0.4, 0.7, 1.4, 2.0 Nuevo Méjico | 0.1, 0.2, 0.4, 0.7, 1.4, 2.0 Nuevo Méjico | 0.1, 0.2, 0.4, 0.7, 1.4, 2.0 Nuevo Méjico |
| Precisión de longitud de onda | ±0.15 nm | ±0.20 nm | ±0.15 nm |
| Repetibilidad de longitud de onda | ≤0.05 nm | ≤0.10 nm | ≤0.05 nm |
| Spectral bandwidth deviation | ±0.02 nm | ±0.02 nm | ±0.02 nm |
| Lamp stand | 8 lamps; simultaneous preheating of 1–4 lamps | 8 lamps; simultaneous preheating of 1–4 lamps | 8 lamps; simultaneous preheating of 1–4 lamps |

Rendimiento fotométrico
| Parámetro | AA1800 | AA1900 | AA2000 |
|---|---|---|---|
| Reading modes | Transmitancia, absorbancia, concentración | Transmitancia, absorbancia, concentración | Transmitancia, absorbancia, concentración |
| rango fotométrico | 0–125%T; −0.1–3.00 Abs | 0–125%T; −0.1–3.00 Abs | 0–125%T; −0.1–3.00 Abs |
| Baseline stability | ±0.002 Abs (steady 30 mín.); ±0.004 Abs (dynamic 15 mín.) | ±0.003 Abs (steady 30 mín.); ±0.005 Abs (dynamic 15 mín.) | ±0.002 Abs (steady 30 mín.); ±0.004 Abs (dynamic 15 mín.) |
| Background correction | D₂ lamp ≥40-fold (1 abdominales); self-reversal ≥80-fold (1 abdominales) | D₂ lamp ≥40-fold (1 abdominales) | D₂ lamp ≥40-fold (1 abdominales); self-reversal ≥80-fold (1 abdominales) |
Flame analysis
| Parámetro | AA1800 | AA1900 | AA2000 |
|---|---|---|---|
| Characteristic concentration (Cu) | ≤0.02 μg/mL | ≤0.035 μg/mL | ≤0.02 μg/mL |
| Detection limit (Cu) | ≤0.004 μg/mL | ≤0.006 μg/mL | ≤0.004 μg/mL |
| Repetibilidad (Cu) | ≤0.5% | ≤0.6% | ≤0.5% |
| Quemador | Titanium burner head | Titanium burner head | Titanium burner head |
| Nebulizer | High-efficiency glass nebulizer | High-efficiency glass nebulizer | High-efficiency glass nebulizer |
| Seguridad | Flame gas & auxiliary gas abnormal-pressure protection | Flame gas & auxiliary gas abnormal-pressure protection | Flame gas & auxiliary gas abnormal-pressure protection |
Titanium burner and nebulizer
AA1800 uses a titanium burner head with a high-efficiency glass nebulizer. Flame gas and auxiliary gas abnormal-pressure protection is standard.
Data processing
| Parámetro | AA1800 | AA1900 | AA2000 |
|---|---|---|---|
| Concentration calculation | Standard curve (6 linear/non-linear fitting methods), standard addition and interpolation | Standard curve (6 linear/non-linear fitting methods), standard addition and interpolation | Standard curve (6 linear/non-linear fitting methods), standard addition and interpolation |
| Repeated measurements | 1–30 times; average, SD and RSD of absorbance/concentration | 1–30 times; average, SD and RSD of absorbance/concentration | 1–30 times; average, SD and RSD of absorbance/concentration |
Accessories and physical data
| Parámetro | AA1800 | AA1900 | AA2000 |
|---|---|---|---|
| Standard accessories | Cu lamp ×1; oil-free air compressor; software; USB flash drive | Cu lamp ×1; Cd lamp ×1; oil-free air compressor; software; USB flash drive; graphite tube ×5; pipette ×1 (5–50 μL); pipette tips ×1000 (200 μL) | Cu lamp ×1; Cd lamp ×1; oil-free air compressor; software; USB flash drive; graphite tube ×5; pipette ×1 (5–50 μL); pipette tips ×1000 (200 μL) |
| Optional accessories | Oxygen-enriched high-temperature flame; hydride generator; audit-trail software; flame autosampler | Oxygen-enriched high-temperature flame; hydride generator; audit-trail software; recirculating chiller; graphite-furnace autosampler; flame autosampler | Oxygen-enriched high-temperature flame; hydride generator; audit-trail software; recirculating chiller; graphite-furnace autosampler; flame autosampler |
| Power supply | C.A. 220 V, 50/60 Hz | C.A. 220 V, 50/60 Hz | C.A. 220 V, 50/60 Hz |
| External size (W×D×H) | 830×650×560 mm | 830×650×560 mm | 830×650×560 mm |
| Package size (W×D×H) | 1010×780×830 mm | 1010×780×830 mm | 1010×780×830 mm |
| Net weight | 85 kilos | 120 kilos | 125 kilos |
| Gross weight | 135 kilos | 183 kilos | 183 kilos |
Service connections and installation

Rear-panel mains inlet, gas and water connectors are grouped for bench installation. The instrument sits on vibration-isolation feet; unpacking and leveling are required before atomizer alignment.
Frequently asked questions
What elements can this AAS determine?
The series is specified for trace and major element determination in environmental monitoring, food safety testing and metallurgical analysis. Element-specific detection limits are supplied for Cu and Cd in the datasheet; limits for other elements must be confirmed against your application.
What is the wavelength range?
All three models cover 185–900 nm with a 1800 lines/mm grating blazed at 250 Nuevo Méjico.
How many lamps can be fitted?
The turret holds eight lamps and can preheat 1–4 lamps simultaneously.
Does AA1800 include a graphite furnace?
No. AA1800 is a dedicated flame system. For graphite furnace atomization, specify AA1900 or AA2000.
Which background correction is fitted?
AA1800 and AA2000 specify D₂ lamp ≥40-fold (1 abdominales) and self-reversal ≥80-fold (1 abdominales). AA1900 specifies D₂ lamp ≥40-fold (1 abdominales).
What data processing is available?
Standard curve with six linear/non-linear fitting methods, standard addition and interpolation; 1–30 repeated measurements with average, SD and RSD.
What power supply is required?
C.A. 220 V, 50/60 Hz for all three models.
Related models
Compare also: BRT AA1900 BRT AA2000.
See the complete BRT AAS series comparison.