GTL-50 Steel Structure Fire Retardant Coating Sample Test Furnace strictly meets OIN 834, BS 476 (Parties 20/21), DANS 13381-4/8, ASTM E119, UL 263, et FR 14907-2018 normes. Features touch screen & wireless PC dual control, diesel spray heating up to 1210°C / 360 min, constant 20±3Pa furnace pressure control, auto thermocouple fault elimination, and automatic curve logging for steel coating fire resistance evaluation.

UGS: GTL-50, Catégories: ,
Demande maintenant

ADRESSE EMAIL

PRÉNOM

NOM DE FAMILLE

WhatsApp

MESSAGE

Le GTL-50 Steel Structure Fire Retardant Coating Sample Test Furnace is a dedicated laboratory testing instrument developed strictly in accordance with OIN 834, BS 476 (Parties 20/21), DANS 13381-4/8, ASTM E119, UL 263, et FR 14907-2018 normes. It is engineered to evaluate the fire resistance rating, thermal insulation efficiency, and flame retardancy of fireproof coatings applied to steel plates and structural components.

Featuring a compact 1.6m × 1.6m × 1.6m diesel-fueled heating chamber, dual touch screen and wireless PC control, real-time closed-loop temperature and pressure curve tracking (up to 1210°C / 360 min), and automatic thermocouple fault elimination, cette haute précision Fire-Retardant Coating Thermal Insulation Efficiency Test Furnace delivers authoritative test data for paint manufacturers, third-party fire testing laboratories, marine classification bodies, and construction material quality supervision bureaus.


Steel Structure Fire Retardant Coating Sample Test Furnace Compliance & Normes

This fire-retardant coating sample test furnace enables certified testing facilities to assess coating performance across global structural fire protection frameworks:

Code standard Titre standard & Portée Testing Application
OIN 834 / BS 476-20/21 Fire-resistance tests — Elements of building construction & loadbearing structural elements International benchmark for standard time-temperature fire curve compliance and coating testing
DANS 13381-8 / DANS 13381-4 Test methods for determining the contribution to the fire resistance of structural members (Reactive & Passive protection to steel) European standard for intumescent (reactive) and thick non-intumescent (passive) steel fireproof coatings
ASTM E119 / UL 263 Standard Test Methods for Fire Tests of Building Construction and Materials North American fire resistance and thermal insulation evaluations for structural steel coatings
FR 14907-2018 & GB/T 9978-2008 Fire-retardant coating for steel structures & Fire-resistance tests for building elements Mandatory Chinese national standard for thermal insulation efficiency & fire resistance limit testing
IMO FTP Code Part 11 International Code for Application of Fire Test Procedures — Structural Fire Protection Fire protection coating evaluations for marine steel structures, deck bulkheads, and offshore platforms

Technological Features of GTL-50 Fire-Retardant Coating Test Furnace

  • Dual Touch Screen & Wireless PC Multi-Station Linkage: Operates via an intuitive standalone onboard touch screen or wireless PC software. Supports multi-furnace synchronization from a single computer terminal and allows centralized data monitoring across network workstations.
  • High-Efficiency Diesel Spray Burner & Standard Curve Simulation: Equipped with an automatic diesel combustion system (using 0# / -10# diesel fuel) capable of following standard time-temperature curves up to 1210°C over a 360-minute duration (par ex., 60 min @ 925°C, 180 min @ 1115°C, 360 min @ 1210°C).
  • Automated Closed-Loop Pressure & Temperature Regulation: Maintains a constant furnace pressure of 20 ± 3 Pa (0–100 Pa measurement range) and a tight unexposed surface temperature tolerance within ±4°C, vastly surpassing national standard precision thresholds.
  • Intelligent Thermocouple Fault Elimination & Flame Protection: Built-in self-diagnostics automatically detect, isolate, and shield damaged thermocouple signals during continuous heating. Features automatic re-ignition protection in the event of accidental flameout during testing.
  • Automated Post-Test Exhaust Cooling & Power-Off Protection: Includes forced exhaust heat dissipation upon test completion to accelerate turn-around times. Built-in power-off backup programs prevent data loss and equipment damage during sudden electrical interruptions.

Technical Specifications of Steel Structure Fire Retardant Coating Test Furnace

Normes de test régissant OIN 834, BS 476-20/21, DANS 13381-4/8, ASTM E119, UL 263, FR 14907-2018, GB/T 9978-2008, IMO FTP Code Part 11
Standard Specimen Dimensions 500 mm × 500 mm × 6 mm (Coated steel plate)
Heating Fuel & System 0# / -10# Diesel; full-automatic spray burner assembly
Programmed Temperature Curves 60 min: 925°C | 120 min: 1050°C | 180 min: 1115°C | 240 min: 1150°C | 300 min: 1180°C | 360 min: 1210°C
Furnace & Surface Thermocouples 4 furnace sensors + 2 unexposed surface sensors (Class II K-type sheathed); Tolerance: ±4°C
Furnace Pressure Control Range: 0–100 Pa; Closed-loop constant control: 20 ± 3 Pa
Airflow Volume & Précision 0–50 m³/min; Measurement accuracy: ≤ ±0.5 m³/min
Timer & Data Storage Timing range: 0–360 min (< ±2s accuracy); Onboard memory stores >100 test groups
Alimentation & Consumption AC 220V ±10%, 50 Hz; Total Power: 3.0 kW
Overall Equipment Dimensions 1.6 m (L) × 1.6 m (W) × 1.6 m (H)

System Architecture & Testing Workflow of Fire Retardant Coating Test Furnace

Flux de travail de test standardisé:

  1. Apply the steel structure fire-retardant coating to the 500mm × 500mm × 6mm standard steel plate specimen and allow it to cure per standard instructions.
  2. Mount the specimen onto the test furnace opening, attaching K-type thermocouples to the unexposed face to measure heat conduction.
  3. Select the target fire-resistance duration (par ex., 180 ou 360 minutes) on the touch screen or wireless PC interface.
  4. Trigger the automatic spray burner; the system heats up following the GB 14907 / OIN 834 curve while continuously recording furnace pressure and unexposed surface temperature.
  5. Upon test completion, the furnace automatically terminates heating, saves test records, and initiates forced exhaust cooling.

Applications client cibles:

  • Thin, ultra-thin, and thick steel structure fireproof coating manufacturers.
  • Third-party fire protection product inspection centers and construction quality laboratories.
  • Chemical engineering and polymer R&D institutes specializing in intumescent flame-retardant coatings.
  • University civil engineering and fire safety science research centers.

Steel Structure Fire Retardant Coating Sample Test Furnace FAQ

T1: What is the main evaluation metric for GB 14907 steel structure coating tests?

UN: The primary metric is the Thermal Insulation Efficiency / Fire Resistance Limit, defined as the time required for the average temperature of the unexposed steel plate to reach a critical limit (typically 500°C or specified temperature rise) under the standard fire time-temperature curve.

T2: How does the wireless PC control benefit multi-furnace laboratory operations?

UN: Wireless linkage enables a single central computer terminal to wirelessly connect to and synchronously control multiple GTL-50 test furnaces. Laboratories can execute simultaneous batch tests, collect live data, and archive reports without extensive cabling.

T3: What safety mechanisms are built into the diesel combustion system?

UN: The GTL-50 includes automatic flameout detection with auto re-ignition capabilities, power-off memory protection, and over-temperature safety cut-offs to guarantee safe, unattended laboratory testing.

Test Parameter

Flame Spread

Standard

BS 476, FR 14907-2018, GB/T 9978-2008, OIN 834

Avis

Il n'y a pas encore d'avis.

Soyez le premier à donner votre avis "GTL-50 Steel Structure Fire Retardant Coating Sample Test Furnace”

Votre adresse email ne sera pas publiée. Les champs obligatoires sont marqués *