O Testador de inflamabilidade vertical têxtil FCZ-50 is a professional laboratory testing system designed to evaluate the flame resistance, afterflame time, afterglow time, and damaged length (char length) of vertically oriented textiles and protective fabrics. Built in compliance with ASTM D6413, 16 CFR Part 1615/1616, ISO 6940/6941, BS EN ISO 15025, e padrão nacional GB/T 5455-2014, this apparatus provides standardized test conditions for flame-retardant fabric qualification.
Featuring a precision 11 mm inner-diameter Bunsen burner, dual specimen conditioning modes (3 s ignition for oven-dried samples and 12 s ignition for standard conditioned samples), a full range of standardized char length tear weights (54.5 g to 453.6 g), and high-resolution timing (0.1 precisão), isso avançado Textile Vertical Burning Tester is an essential tool for garment manufacturers, technical textile mills, aerospace interior suppliers, e laboratórios de testes terceirizados.
Textile Vertical Flammability Tester Compliance & Padrões
This vertical fabric burning test system enables certified laboratories to conduct formal flame resistance and char length evaluations across international and regional standards:
| Código Padrão |
Título padrão & Escopo |
Aplicação de teste primário |
| ASTM D6413 / D6413M |
Standard test method for flame resistance of textiles (vertical test) |
North American benchmark for flame-resistant workwear, protective apparel, and industrial fabrics |
| 16 CFR Part 1615 / 1616 |
Standard for the flammability of children’s sleepwear (Sizes 0–6X & 7–14) |
Mandatory U.S. CPSC safety compliance testing for children’s sleepwear and nightwear fabrics |
| ISO 6940 & ISO 6941 |
Textile fabrics — Burning behaviour — Determination of ease of ignition & flame spread properties |
International standard method for vertically oriented interior drapery, curtains, and apparel fabrics |
| BS EN ISO 15025 / Bobagem 5438 |
Protective clothing — Protection against flame — Method of test for limited flame spread |
European safety qualification for firefighter turn-out gear and industrial welder protective suits |
| GB/T 5455-2014 |
Textiles — Burning behaviour — Determination of damaged length, afterglow time and afterflame time |
Mandatory Chinese national test method for vertical flame retardancy of woven and knitted fabrics |
Technological Features of FCZ-50 Textile Flammability Tester
- Dual Conditioning & Ignition Test Profiles: Fully supports both Condition A (12 s ignition with propane/butane for standard-conditioned fabrics) and Condition B (3 s ignition with ≥97% purity methane for oven-dried specimens).
- Standardized Char Length Tear Weight System: Includes five precision-calibrated weights (54.5 g, 113.4 g, 226.8 g, 340.2 g, e 453.6 g) mapped directly to fabric weight per unit area (≤101 g/m² to ≥650 g/m²) for accurate tear-strength damaged length calculation.
- Precision Bunsen Burner Assembly: Features an 11 mm internal nozzle diameter Bunsen burner ensuring consistent gas flow velocity and stable flame geometry during exposure.
- Enclosed Draft-Free Combustion Chamber: Compact stainless steel chamber (329 milímetros × 329 milímetros × 767 milímetros) insulates the flame from ambient air currents while observing combustion via a clear front window.
- High-Precision Measurement & Timing: Electronic timer measuring afterflame and afterglow durations with 0.1 precisão, combined with 1 mm linear resolution for damaged length reporting.
Technical Specifications of FCZ-50 Textile Flammability Machine
| Padrões de teste aplicáveis |
ASTM D6413, 16 CFR 1615/1616, ISO 6940, ISO 6941, BS EN ISO 15025, GB/T 5455-2014 |
| Dimensões internas da câmara de combustão |
(329 ± 2) milímetros (C) × (329 ± 2) milímetros (D) × (767 ± 2) milímetros (H) |
| Dimensões gerais do equipamento |
632 milímetros (eu) × 332 milímetros (C) × 852 milímetros (H) |
| Specimen Holder & Dimensões |
Holder internal frame: 356 milímetros × 51 milímetros; Specimen size: 300 milímetros × 89 milímetros |
| Ignition Burner Specification |
Queimador de Bunsen com 11 mm internal nozzle diameter |
| Test Gas Options & Ignition Time |
• Condition A: Propane, butane, or mixed gas (12 ignição)
• Condition B: Metano (purity ≥97%) (3 ignição) |
| Included Char Length Weights |
54.5 g, 113.4 g, 226.8 g, 340.2 g, e 453.6 g |
| Measurement Resolution |
Tempo: 0.1 é; Damaged Length: 1 milímetros |
| Fonte de energia & Classificação de potência |
AC 220 V ± 10%, 50 hertz; Consumo de energia: 100 C |
Arquitetura do sistema & Testing Workflow of FCZ-50 Textile Flammability Tester
Fluxo de trabalho de teste padronizado:
- Cut specimens to 300 milímetros × 89 mm and condition according to Condition A (standard atmosphere) or Condition B (oven-dried at 105°C).
- Mount the sample inside the 356 milímetros × 51 mm vertical specimen holder and suspend it in the combustion chamber.
- Select test gas (propane/butane for Condition A or Methane ≥97% for Condition B) and set ignition timer (12 s or 3 é).
- Apply flame exposure. Automatically record afterflame time (duration sample continues burning after burner shuts off) and afterglow time.
- Attach the designated weight (based on fabric mass per unit area from Table 2) to the bottom corner of the burned sample, lift slowly, and measure the resulting damaged length (char length).
Aplicativos de clientes-alvo:
- Protective apparel, firefighter turnout gear, and welder suit manufacturing plants.
- Children’s sleepwear, curtain/drapery, and home textile quality control laboratories.
- Aerospace interior and military fabric certification organizations.
FCZ-50 Textile Vertical Flammability Tester FAQ
1º trimestre: What parameters are measured during a vertical fabric flammability test?
UM: The FCZ-50 evaluates three critical parameters: Afterflame Time (duration of visible flame after burner removal), Afterglow Time (duration of flameless glowing combustion), and Damaged Length (char length measured using standardized tear weights).
2º trimestre: How is the char length weight chosen for testing?
UM: The tear weight is determined by fabric mass per unit area (g/m²):
- ≤ 101 g/m²: 54.5 g weight
- > 101 para ≤ 207 g/m²: 113.4 g weight
- > 207 para ≤ 338 g/m²: 226.8 g weight
- > 338 para < 650 g/m²: 340.2 g weight
- ≥ 650 g/m²: 453.6 g weight
3º trimestre: Can this device be used for ASTM D6413 workwear testing?
UM: Sim. The vertical specimen frame, 11 mm Bunsen burner nozzle, 12 s ignition timer, and specified char length weights align with ASTM D6413 requirements.