BRT AA1900 flame and graphite furnace AAS with longitudinal furnace heating, titanium burner and eight-lamp turret. Complete technical specifications.
- Product overview
- Key features across the series
- Optical system and photometric performance
- الأداء الضوئي
- Flame analysis
- Graphite furnace analysis
- Inside the sample compartment
- Autosampler and front-panel controls
- Data processing
- Accessories and physical data
- Service connections and installation
- Frequently asked questions
- Related models
- Request a quotation
Product overview
ال BRT AA1900 integrates flame and graphite furnace atomization in one system. It is built for trace and major element determination in environmental monitoring, food safety testing and metallurgical analysis, combining a titanium flame burner with longitudinal graphite furnace heating.

Key features across the series
- Total-reflection optical system designed for high signal-to-noise performance.
- Eight-lamp turret with 360° rotation and simultaneous preheating of 1–4 lamps.
- Windows workstation software with automatic control, QA/QC functions, SD/RSD calculation and calibration-curve correlation checking.
- Multiple data-processing modes including average, peak height, peak area and statistical analysis.
- One-click database access for wavelength, عرض النطاق الترددي الطيفي, lamp current, negative high voltage and energy-balance settings.
Optical system and photometric performance
| المعلمة | AA1800 | AA1900 | AA2000 |
|---|---|---|---|
| Instrument type | Flame system | Flame & GFAA integrated | Flame & GFAA integrated |
| Beam type | Double-beam | Double-beam | Double-beam |
| Diffraction grating | 1800 خطوط / مم; blaze wavelength 250 نانومتر | 1800 خطوط / مم; blaze wavelength 250 نانومتر | 1800 خطوط / مم; blaze wavelength 250 نانومتر |
| نطاق الطول الموجي | 185-900 نانومتر | 185-900 نانومتر | 185-900 نانومتر |
| عرض النطاق الترددي الطيفي | 0.1, 0.2, 0.4, 0.7, 1.4, 2.0 نانومتر | 0.1, 0.2, 0.4, 0.7, 1.4, 2.0 نانومتر | 0.1, 0.2, 0.4, 0.7, 1.4, 2.0 نانومتر |
| دقة الطول الموجي | ±0.15 nm | ±0.20 nm | ±0.15 nm |
| تكرار الطول الموجي | ≤0.05 nm | ≤0.10 nm | ≤0.05 nm |
| Spectral bandwidth deviation | ±0.02 nm | ±0.02 nm | ±0.02 nm |
| Lamp stand | 8 lamps; simultaneous preheating of 1–4 lamps | 8 lamps; simultaneous preheating of 1–4 lamps | 8 lamps; simultaneous preheating of 1–4 lamps |

الأداء الضوئي
| المعلمة | AA1800 | AA1900 | AA2000 |
|---|---|---|---|
| Reading modes | النفاذية, الامتصاص, concentration | النفاذية, الامتصاص, concentration | النفاذية, الامتصاص, concentration |
| Photometric range | 0–125%T; −0.1–3.00 Abs | 0–125%T; −0.1–3.00 Abs | 0–125%T; −0.1–3.00 Abs |
| Baseline stability | ±0.002 Abs (steady 30 دقيقة); ±0.004 Abs (dynamic 15 دقيقة) | ±0.003 Abs (steady 30 دقيقة); ±0.005 Abs (dynamic 15 دقيقة) | ±0.002 Abs (steady 30 دقيقة); ±0.004 Abs (dynamic 15 دقيقة) |
| Background correction | D₂ lamp ≥40-fold (1 القيمة المطلقة); self-reversal ≥80-fold (1 القيمة المطلقة) | D₂ lamp ≥40-fold (1 القيمة المطلقة) | D₂ lamp ≥40-fold (1 القيمة المطلقة); self-reversal ≥80-fold (1 القيمة المطلقة) |
Flame analysis
| المعلمة | AA1800 | AA1900 | AA2000 |
|---|---|---|---|
| Characteristic concentration (Cu) | ≤0.02 μg/mL | ≤0.035 μg/mL | ≤0.02 μg/mL |
| Detection limit (Cu) | ≤0.004 μg/mL | ≤0.006 μg/mL | ≤0.004 μg/mL |
| التكرار (Cu) | ≤0.5% | ≤0.6% | ≤0.5% |
| الموقد | Titanium burner head | Titanium burner head | Titanium burner head |
| Nebulizer | High-efficiency glass nebulizer | High-efficiency glass nebulizer | High-efficiency glass nebulizer |
| أمان | Flame gas & auxiliary gas abnormal-pressure protection | Flame gas & auxiliary gas abnormal-pressure protection | Flame gas & auxiliary gas abnormal-pressure protection |
Graphite furnace analysis
| المعلمة | AA1800 | AA1900 | AA2000 |
|---|---|---|---|
| Characteristic mass | — | Cd ≤0.6 pg; Cu ≤20 pg | Cd ≤0.5 pg; Cu ≤20 pg |
| Detection limit | — | Cd ≤1.0 pg; Cu ≤25 pg | Cd ≤0.8 pg; Cu ≤25 pg |
| التكرار (Cd) | — | ≤3% | ≤2.5% |
| نطاق درجة الحرارة | — | RT–3000 °C | RT–3000 °C |
| Heating rate | — | Max. 2000 °C/s | Max. 2000 °C/s |
| أمان | — | Cooling-water flow, protective-gas pressure, furnace-body temperature, power-supply temperature and graphite-tube installation alarms | Cooling-water flow, protective-gas pressure, furnace-body temperature, power-supply temperature and graphite-tube installation alarms |
Inside the sample compartment
The graphite furnace workhead, injection assembly and sample bottle sit in the open sample compartment. Longitudinal furnace heating provides uniform gas-phase temperature, rapid atomization and stable analytical performance.

Autosampler and front-panel controls

Data processing
| المعلمة | AA1800 | AA1900 | AA2000 |
|---|---|---|---|
| Concentration calculation | Standard curve (6 linear/non-linear fitting methods), standard addition and interpolation | Standard curve (6 linear/non-linear fitting methods), standard addition and interpolation | Standard curve (6 linear/non-linear fitting methods), standard addition and interpolation |
| Repeated measurements | 1–30 times; average, SD and RSD of absorbance/concentration | 1–30 times; average, SD and RSD of absorbance/concentration | 1–30 times; average, SD and RSD of absorbance/concentration |
Accessories and physical data
| المعلمة | AA1800 | AA1900 | AA2000 |
|---|---|---|---|
| Standard accessories | Cu lamp ×1; oil-free air compressor; برمجة; USB flash drive | Cu lamp ×1; Cd lamp ×1; oil-free air compressor; برمجة; USB flash drive; graphite tube ×5; pipette ×1 (5–50 μL); pipette tips ×1000 (200 μL) | Cu lamp ×1; Cd lamp ×1; oil-free air compressor; برمجة; USB flash drive; graphite tube ×5; pipette ×1 (5–50 μL); pipette tips ×1000 (200 μL) |
| Optional accessories | Oxygen-enriched high-temperature flame; hydride generator; audit-trail software; flame autosampler | Oxygen-enriched high-temperature flame; hydride generator; audit-trail software; recirculating chiller; graphite-furnace autosampler; flame autosampler | Oxygen-enriched high-temperature flame; hydride generator; audit-trail software; recirculating chiller; graphite-furnace autosampler; flame autosampler |
| Power supply | AC 220 V, 50/60 هرتز | AC 220 V, 50/60 هرتز | AC 220 V, 50/60 هرتز |
| External size (W×D×H) | 830×650×560 mm | 830×650×560 mm | 830×650×560 mm |
| Package size (W×D×H) | 1010×780×830 mm | 1010×780×830 mm | 1010×780×830 mm |
| Net weight | 85 كجم | 120 كجم | 125 كجم |
| Gross weight | 135 كجم | 183 كجم | 183 كجم |
Service connections and installation

Rear-panel mains inlet, gas and water connectors are grouped for bench installation. The instrument sits on vibration-isolation feet; unpacking and leveling are required before atomizer alignment.
Frequently asked questions
What elements can this AAS determine?
The series is specified for trace and major element determination in environmental monitoring, food safety testing and metallurgical analysis. Element-specific detection limits are supplied for Cu and Cd in the datasheet; limits for other elements must be confirmed against your application.
What is the wavelength range?
All three models cover 185–900 nm with a 1800 lines/mm grating blazed at 250 نانومتر.
How many lamps can be fitted?
The turret holds eight lamps and can preheat 1–4 lamps simultaneously.
What does the graphite furnace add?
Graphite furnace atomization extends the instrument to lower mass detection limits — the datasheet lists Cd ≤1.0 pg (AA1900) and Cd ≤0.8 pg (AA2000), with a temperature range up to 3000 درجة مئوية.
Which background correction is fitted?
AA1800 and AA2000 specify D₂ lamp ≥40-fold (1 القيمة المطلقة) and self-reversal ≥80-fold (1 القيمة المطلقة). AA1900 specifies D₂ lamp ≥40-fold (1 القيمة المطلقة).
What data processing is available?
Standard curve with six linear/non-linear fitting methods, standard addition and interpolation; 1–30 repeated measurements with average, SD and RSD.
What power supply is required?
AC 220 V, 50/60 Hz for all three models.
Related models
Compare also: BRT AA1800 BRT AA2000.
See the complete BRT AAS series comparison.
Request a quotation
Specify your element list, sample matrix, required detection limits and atomization mode (flame only or flame + graphite furnace). BRT will confirm the configuration, accessory set and delivery terms. Contact info@brt-test.com or WhatsApp +86 156 0405 7553.