BRT-5600UV UV Vis Spectrophotometer is a 17 kg benchtop split double-beam instrument featuring a 320×240 blue LCD, 190–1100 nm spectral range, 1.8 nm bandwidth, C-T monochromator with 1200 lines/mm grating, and an auto 8-cell holder. Delivering ±0.3 nm wavelength accuracy, ≤0.05%T stray light, and ≤0.001 Abs/30 min baseline drift, it offers 200-dataset memory, USB connectivity, and LPT printer support for precise analytical testing.

SKU: BRT-5600UV, Categories: ,
Inquiry Now

EMAIL ADDRESS

FIRST NAME

LAST NAME

whatsapp

MESSAGE

The BRT-5600UV UV Vis Spectrophotometer is an ultra-precision, split double-beam analytical instrument built on a Czerny-Turner (C-T) monochromator optical architecture, a 1200 lines/mm holographic grating, and an automated 8-cell holder system. Designed for high-accuracy analytical laboratories, pharmaceutical quality assurance, environmental testing agencies, and life science research facilities, it delivers exceptional baseline stability, low stray light (≤0.05%T), and minimal baseline drift (≤0.001 Abs/30 min) across the 190 nm to 1100 nm spectral range with a 1.8 nm bandwidth, ±0.3 nm wavelength accuracy, and 0.2 nm wavelength repeatability.

Featuring dual deuterium and tungsten halogen light sources with adjustable switching points (294–365 nm), this advanced UV Vis Spectrophotometer guarantees high photometric accuracy (±0.3%T) and reliable quantitative concentration linearity. Equipped with a large 320×240 blue LCD display, automatic startup wavelength calibration, onboard memory for 200 datasets and standard calibration curves, USB data export, and LPT printer compatibility, the 17 kg BRT-5600UV delivers reliable, batch-driven laboratory analysis.


UV Vis Spectrophotometer Measurement & Analysis Capabilities

This automated split double-beam monochromator spectrophotometer equips technicians to perform high-precision photometric, quantitative, and kinetic testing workflows across liquid and solution samples:

Analysis Mode Methodology & Operational Scope Primary Testing Application
Photometric Measurement Direct evaluation of Absorbance (Abs) and Transmittance (%T) at user-selected fixed wavelengths across 190–1100 nm with 0.1 nm resolution High-precision sample optical density screening, raw material verification, and solution clarity testing
Automated Quantitative Concentration Profiling Batch concentration determinations using factor mode ($C = KA + B$) or multi-point standard curve fitting combined with the automatic 8-cell changer Pharmaceutical active ingredient assays, environmental water pollutant checks (COD, nitrate, heavy metals), and chemical quality control
PC-Driven Kinetics & Spectral Analysis Time-course reaction tracking, dynamic kinetics, and multi-wavelength measurements using optional professional PC software via USB connection Enzymatic reaction dynamics, nucleic acid/protein assays, dynamic degradation monitoring, and complex chemical profiling

Technological Features of BRT-5600UV UV Vis Spectrophotometer

  • Split Double-Beam Optics & C-T Monochromator

    Combines a split double-beam optical architecture with a Czerny-Turner monochromator and 1200 lines/mm holographic grating to deliver ultra-low stray light (≤0.05%T) and ±0.3 nm wavelength accuracy.

  • Automated 8-Position Cell Holder

    Features an integrated automatic 8-cell holder system that streamlines multi-sample processing, boosting batch analysis efficiency and minimizing manual handling errors.

  • Ultra-Low Baseline Drift & High Stability

    Engineered for continuous operation with an optical baseline drift of ≤0.001 Abs per half hour, preserving test integrity during extended analytical runs.

  • Dual Light Source System with Customizable Switching

    Integrates an imported deuterium lamp and a 20W tungsten halogen lamp with customizable light source switching points between 294 nm and 365 nm to optimize lamp performance and longevity.

  • 320×240 Blue LCD Display & Onboard Memory

    Provides a clear 320×240 blue LCD screen interface displaying simultaneous %T and Abs values, with onboard memory capable of storing up to 200 datasets and standard curves.

  • Dual Output Ports (USB PC Interface & LPT Parallel Port)

    Equipped with a USB interface for data transfer to PC analysis software and an LPT parallel port for outputting measurement results to dedicated laboratory printers.


Technical Specifications of UV Vis Spectrophotometer

Product Model BRT-5600UV Low-Drift Double Beam UV Vis Spectrophotometer
Optical System & Grating Split Double Beam C-T Monochromator, 1200 lines/mm Holographic Grating
Wavelength Spectrum & Bandwidth 190 nm to 1100 nm; Spectral Bandwidth: 1.8 nm; Wavelength Resolution: 0.1 nm
Wavelength Accuracy & Repeatability Accuracy: ±0.3 nm; Repeatability: ≤0.2 nm
Photometric Range & Accuracy 0 to 200%T, -0.301 to 3.000 Abs, -9999 to 9999 C; Transmittance Accuracy: ±0.3%T
Optical Stability & Stray Light Stray Light: ≤0.05%T; Baseline Drift: ≤0.001 Abs/30 min
Light Sources & Sample Holder 20W Tungsten Halogen Lamp & Deuterium Lamp; Automatic 8-Position Cell Holder
Display & Data Storage 320 × 240 Blue LCD Screen; Stores 200 Datasets & Standard Calibration Curves
Dimensions & Instrument Weight 545 × 450 × 235 mm (L × W × H); Net Weight: 17 kg (37.48 lb); Gross Weight: 26 kg

Standard Package Accessories Included

  • Main Host Unit: BRT-5600UV UV Vis Spectrophotometer (1 unit).
  • Sample Cell Holder: Automatic 8-Position Cell Holder (installed).
  • Quartz Cuvettes: 10 mm Standard Optical Path Quartz Cuvettes (4 pcs for Ultraviolet spectrum testing).
  • Glass Cuvettes: 10 mm Standard Optical Path Glass Cuvettes (4 pcs for Visible spectrum testing).
  • Power & Connectivity: Dual-Voltage (110V/220V) AC Power Cable, USB Interface Cable.
  • Documentation & Accessories: User Manual, Quality Certificate, Warranty Card.

Testing Workflow & Target Applications

Standardized Testing Workflow:

  1. Power on the BRT-5600UV; allow the microcomputer to execute lamp ignition, zeroing, and startup wavelength self-calibration.
  2. Select Photometric or Quantitative mode on the 320×240 blue LCD display screen.
  3. Load a blank reference solvent into position 1 of the automated 8-cell changer and execute auto 0A / 100%T baseline zeroing.
  4. Load up to 7 test samples into the remaining changer slots to automatically perform sequential quantitative concentration analysis.
  5. Store measurement datasets (up to 200 sets) in local memory, transmit via USB interface to optional PC analysis software, or print reports via the LPT parallel printer port.

Target Customer Applications:

  • Pharmaceutical testing facilities, active pharmaceutical ingredient (API) labs, and QA/QC centers.
  • Environmental monitoring agencies (nitrate, phosphate, COD, and heavy metal assays).
  • Biochemistry laboratories, life science research institutes, and university research facilities.
  • Chemical processing plants, food safety inspection stations, and agriculture analytical laboratories.

UV Vis Spectrophotometer FAQ

Q1: What are the main benefits of the ≤0.001 Abs/30 min baseline drift in the BRT-5600UV?

A: Ultra-low baseline drift ensures that the instrument maintains optical baseline flatness over long testing shifts. This eliminates frequent manual recalibrations and guarantees high quantitative repeatability.

Q2: How does the ±0.3 nm wavelength accuracy improve research results?

A: The high wavelength accuracy of ±0.3 nm, combined with 0.2 nm repeatability, ensures precise alignment with target absorption peaks. This minimizes measurement variance across complex chemical and biological assays.

Q3: Can the light source switching point be adjusted manually?

A: Yes, the light source switching point between the deuterium lamp and the tungsten halogen lamp can be adjusted within the range of 294 nm to 365 nm to accommodate specific sample absorption characteristics.

Reviews

There are no reviews yet.

Be the first to review “BRT-5600UV UV Vis Spectrophotometer”

Your email address will not be published. Required fields are marked *